Tu trouveras les réponses dans ce document
On a lu toutes les pages de 66 pour que tu n'aies pas à le faire. Pose ta question et obtiens la réponse tout de suite.
Micro Tymp 2 Tympanometric System 23640 handle, 71170 printer charger (Includes CE Model)
Welch\Allyn
Service Manual
Micro Tymp 2
Tympanometric System
23640 handle, 71170 printer charger
(Includes CE Model)
4341 State Street Road
P.O. Box 220
Skaneateles Falls, NY 13153-0220
Copyright 1995
| Part | Rev | Desc. of Change | ECN # | Date | Initiator |
| 236409 | B | New Release | 5-28225 | 6/95 | R. Settembre |
| 236409 | C | Added CE Updates & Cautions | 5-35489 | 6/97 | R. Settembre |
NOTE: Drawings & illustrations in this document are for reference only. Consult system or call factory for latest revisions.

MiroOmn2 Wkth Abyn
Micro Tymp 2 Tympanometric System
23640 handle, 71170 printer charger (Includes CE Model)
Copyright 1995 Welch Allyn Co.
Welch Allyn
FORWARD
This manual provides technical assistance for calibrating, diagnosing, and servicing the Welch Allyn MICROTYMP 2 handle and Printer/Charger.
The MicroTymp 2 and or Printer Charger will be returned to the Technical Service Department for either Recalibration (handle only) or for diagnosis and repair. For Printer Charger tests, refer to current factory approved diagnostic tests. Once the fault has been determined, refer to the Manual chapters I and J for the appropriate Removal and Replacement sequence. Re-test the Printer Charger after the repair is completed. Similarly, when servicing the MicroTymp 2 handle, refer to the most recent troubleshooting guide and Calvin Users guide. Once the failed component has been identified, refer to the Manual chapters B through H for the appropriate Removal and Replacement sequence. Re-test (re-calibrate) the handle once the repair has been completed.
The repair ‘TASK’ procedures (R & R .. Removal and Replacement) are written to provide quick and easy reference for different experience levels of repair technician. Required specifications and materials are listed at each step where applicable. For example:
Instructions will start off with a “TASK, TO DO.” For example, B. 3. “Unplug Flex Circuit” statement. This will be followed by a “TECHNIQUE, HOW TO DO” statement. Follow these steps to preform the task.— “Grasp pump assembly flex circuit with pin straightening pliers.”
This manual is written in a style useful for training and/or translation. Set up training sessions to allow the trainee to practice the use of the diagnostic, repair, adjustment, and re-testing procedures found in this manual. Set up hands-on practice sessions to enable the technician to develop and retain skills. Provide each trainee with the opportunity to practice the procedure(s) after it has been demonstrated by the instructor.
IMPORTANT:
Welch Allyn service part numbers and material numbers appearing in this manual are for the purpose of familiarizing new technicians with the Welch Allyn parts numbering system. Order replacement parts, as always, by referencing your latest bill of materials, parts catalog and assembly drawings. Part number changes, product updates, or new test procedures should be noted on the appropriate page of this manual by the manual owner.
SAFE SERVICE
CLEAN AND DISINFECT per instructions in MicroTymp2 Operating Instructions PN 236311 PAGE C-50 prior to repairing, or calibrating all returned MicroTymp 2's.
PERFORM appropriate “A” tests after servicing to assure safe, accurate and reliable product operation.
QUALITY SERVICE
WEAR an ESD wrist strap when servicing the MicroTymp 2 or handling printed circuit boards.
STORE new and used boards in anti-static bags.
UPDATE your Repair & Training Manual by notating changes and revisions of part numbers and procedures.
READ AND UNDERSTAND customer comments and complaints on accompanying documentation (P.O. or other correspondence).
If an injury, illness to patient or user, or hazardous malfunction has occurred or is suspected, PROCEED NO FURTHER! DO NOT disassemble, repair, adjust, or in any way alter the MicroTymp 2 or the printer charger base ... instead, notify your supervisor at once. Give them all accompanying paperwork and any information you might have on the return. They will take appropriate action according to Welch Allyn policy. Check customer service records to determine if this MicroTymp 2 requires or has received any applicable retrofit modifications, improvements, or inspections, and perform these if required.
INSPECT AND TEST it before and after it is repaired. Documentation should include: general condition, defects, missing components and labels, electrical, functional and mechanical problems. Operate the returned instrument to try to duplicate the customer's complaint before and after it is repaired.
TABLE OF CONTENTS / MICRO TYMP 2
Forward
Safe Service
Section 1
Removal and replacement (r & r):
A Normal maintenance
B Power supply board assembly
C Motor/pump assembly
D MCU Board assembly
E Keypad assembly
F LCD Assembly
G ESD Shields
H Valve
I Ballast
J Printer / charger keypad
K Printer and or PCB
Section 2 Preliminary Handle Bench Test
Section 3 Micro Tymp 2 Noise Test Procedure
Section 4 Technical Troubleshooting Guide for Micro Tymp 2 Handles & Printer Chargers
Section 5 Customer Service Questions/Answers
Section 6 CALVIN, Micro Tymp 2 Calibration System User Manual
Section 7 Micro Tymp 2 NEEDS CAL Error Troubleshooting Guide
Section 8 Tools/Test Equipment listing Calvin Components
Section 9 Repair Parts
Section 10 Drawings
Section 11 A-02727 Process Sheet excerpts for 23641 Micro Tymp 2 Handle tests
Section 12 A-02728 Process Sheet excerpts for 71171, 71172, 71174, 71176, 71179 Micro Tymp 2 Printer Charger Tests
A NORMAL MAINTENANCE