Aller au contenu principal

Repair information for blood pressure monitors (also known as sphygmomanometers).

8 questions Voir tout

I used 12v adaptor instead of 6v, now it has stopped working.

Hi,

I used wrong 12v adapter instead of 6v in bp monitor, now it has stopped working. What to do now?

Répondu ! Afficher la réponse J'ai le même problème

Cette question est-elle utile ?

Indice 1
1 commentaire

How can I fix my blood pressure monitor

par

Ajouter un commentaire

2 Réponses

Solution retenue

Hi,

You most probably have a damaged voltage regulator and / or other input supply voltage components on the monitor's control board.

You will have to open it up and inspect it for any obvious damage.

Post some close up pictures of any damaged areas on the control board (hopefully it will be obvious but perhaps it won't be) and maybe someone here will be able to help.

Here's how to do it.

Ajout d'images à une question existante

Cette réponse est-elle utile ?

Indice 6
Ajouter un commentaire

It’s possible it used battery power during the time it worked and due to not getting a charge through the 6v power supply, it’s now drained. Try getting the right power supply and try again.

Cette réponse est-elle utile ?

Indice 2
Ajouter un commentaire

Ajouter une réponse