  1. Airox Legendair Turbine Box Replacement, How to Properly Clean the Airox Legendair: étape 1, image 1 de 1
    • First, switch off the device.

    • Place the device in a clean and open environment.

  2. Airox Legendair Turbine Box Replacement: étape 3, image 1 de 1
    • Using a cloth or a sponge slightly dampened with a bactericide or germicide solution, gently clean the outside of the device.

  3. Airox Legendair Turbine Box Replacement: étape 4, image 1 de 1
    • If necessary, spray the device lightly with the germicide solution, taking account of the requirements discussed in the preliminary operations.

    • Do not allow any liquid to enter inside the device (in particular through the cooling apertures).

    • Use a gentle air blast to clear away any remaining liquid from ventilator.

  4. Airox Legendair Turbine Box Replacement, Turbine Box: étape 6, image 1 de 1
    • Make sure to turn off the ventilator before starting the repair.

    • Put the device in a clean and open environment.

    • Thoroughly clean the device before beginning any work.

  5. Airox Legendair Turbine Box Replacement: étape 7, image 1 de 1
    • Open the device by removing the five attachment screws (Pl1.20) on the bottom casing side (Pl1-280). Lift off the top outer panel.

    • Carefully separate the two casings to avoid tearing off the keypad connections.

  6. Airox Legendair Turbine Box Replacement: étape 8, image 1 de 1
    • Unscrew the three attachment screws (Pl1-130) for the CPU board (Pl1-120).

  7. Airox Legendair Turbine Box Replacement: étape 9, image 1 de 1
    • Disconnect all connectors Pl1.J1.CT, Pl1.J2.CT of the turbine control board.

  8. Airox Legendair Turbine Box Replacement: étape 10, image 1 de 1
    • Extract the turbine casing (Pl1.50) from its housing and uncouple the flow rate measurement block (Pl1-260), the O2 supply pipes and the piezo solenoid valve.

  9. Airox Legendair Turbine Box Replacement: étape 11, image 1 de 1
    • Place the turbine box (Pl1-50) in a clean and open environment

    • Remove the four screws, washers and spacers (Pl1.70, Pl1.80, Pl1.90) in order to successfully extract the turbine control board (Pl1.60).

  10. Airox Legendair Turbine Box Replacement: étape 12, image 1 de 1
    • Disconnect the box from the turbine control board (Pl1.J6.CT).

  11. Airox Legendair Turbine Box Replacement: étape 13, image 1 de 1
    • Replace the turbine box (Pl1-50).

    • Replace the turbine control board (Pl1.60) on the box with four screws, washers and spacers (Pl1.70, Pl1.80, Pl1.90) (Torque 5 Nm max.).

  12. Airox Legendair Turbine Box Replacement: étape 14, image 1 de 2 Airox Legendair Turbine Box Replacement: étape 14, image 2 de 2
    • Connect the box to the turbine control board (Pl1.J6.CT).

    • Couple the flow rate measurement block (Pl1-260) to the casing and connect the two O2 supply pipes and the piezo solenoid valve.

    • Then, reinsert the assembly into its housing. Put the cone support (Pl1-50) into place. Put the additional cone support (Pl1-220) also into place.

  13. Airox Legendair Turbine Box Replacement: étape 15, image 1 de 1
    • Reconnect all connectors (Pl1.J1.CT, Pl1.J2.CT) to the turbine control board.

  14. Airox Legendair Turbine Box Replacement: étape 16, image 1 de 1
    • Reinstall the CPU board (Pl1-120) with the three attachment screws (Pl1-130) (Torque 5 Nm max.)

    • Close the device again, taking all necessary precautions (see Step 2-3)

Conclusion

Perform a final inspection of the device before putting back into service.

Une autre personne a terminé cette réparation.

Noah Clark

Membre depuis le 04/09/20

243 Réputation

1 tutoriel rédigé

Équipe

Cal Poly, Team S7-G15, Paton Spring 2020 Membre de l'équipe Cal Poly, Team S7-G15, Paton Spring 2020

CPSU-PATON-S20S7G15

3 membres

4 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 0

30 derniers jours : 1

Total : 248