Aller au contenu principal

Autosampler Programming (For Old & New Versions)

Ce dont vous avez besoin

  1. Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: étape 1, image 1 de 2 Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: étape 1, image 2 de 2
    • 1. Switch battery "ON".

    • 2. Autosampler: Push "ON" button.

    • It will say ...STANDBY...

  2. Autosampler Programming  (For Old & New Versions), Select Program Mode: étape 2, image 1 de 2 Autosampler Programming  (For Old & New Versions), Select Program Mode: étape 2, image 2 de 2
    • 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM” button to select "PROGRAM".

    • 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.

  3. Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: étape 3, image 1 de 3 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: étape 3, image 2 de 3 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: étape 3, image 3 de 3
    • 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button.

    • 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button.

    • 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.

  4. Autosampler Programming  (For Old & New Versions), Starting Sampling: étape 4, image 1 de 3 Autosampler Programming  (For Old & New Versions), Starting Sampling: étape 4, image 2 de 3 Autosampler Programming  (For Old & New Versions), Starting Sampling: étape 4, image 3 de 3
    • 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED".

    • 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1".

    • 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).

    • It will take some time before it starts sampling.

    • Eventually screen will show: " BOTTLE 1 AFTER 20 PULSES".

Ligne d'arrivée

Une autre personne a terminé cette réparation.

Brooke Mason

Membre depuis le 05/29/19

2 928 Réputation

17 tutoriels rédigés

Équipe

Open-Storm Membre de l'équipe Open-Storm

Community

14 membres

24 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 0

30 derniers jours : 4

Total : 182