Bear 1000 Air and O2 Inlet Manifold Replacement

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. Bear 1000 Air and O2 Inlet Manifold Replacement, Air and O2 Blender: étape 1, image 1 de 1
    • Remove the ribbon cable at the connection on the oxygen blender control board.

  2. Bear 1000 Air and O2 Inlet Manifold Replacement: étape 2, image 1 de 1
    • Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.

  3. Bear 1000 Air and O2 Inlet Manifold Replacement: étape 3, image 1 de 1
    • The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.

  4. Bear 1000 Air and O2 Inlet Manifold Replacement, Air and O2 Inlet Manifold: étape 4, image 1 de 1
    • Remove the pneumatic and electrical connections labeled.

    • Remove the water trap bowls on the rear of the unit.

  5. Bear 1000 Air and O2 Inlet Manifold Replacement: étape 6, image 1 de 1
    • Remove the four screws and remove the manifold.

    • Remove the two screws holding the power supply to the manifold.

Conclusion

To reassemble your device, follow these instructions in reverse order. Calibrate and test the system to see if the replacement was successful

William Burns

Membre depuis le 04/06/20

285 Réputation

2 tutoriels rédigés

Équipe

Cal Poly, Team S7-G12, Paton Fall 2020 Membre de l'équipe Cal Poly, Team S7-G12, Paton Fall 2020

CPSU-PATON-F20S7G12

2 membres

6 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 0

30 derniers jours : 0

Total : 104