Aller au contenu principal

GE Carescape R860 Display Unit Disassembly

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. GE Carescape R860 Display Unit Disassembly, GE Carescape R860 Display Unit Disassembly: étape 1, image 1 de 1
    • Using an opening tool, gently pry the alarm light lens (boxed) from the display unit.

  2. GE Carescape R860 Display Unit Disassembly: étape 2, image 1 de 1
    • Loosen the seven screws that hold the rear housing using a Phillips #0 screwdriver.

    • Remove the rear housing.

  3. GE Carescape R860 Display Unit Disassembly: étape 3, image 1 de 1
    • Cut the cable tie holding the two cables at the lower right side using scissors.

  4. GE Carescape R860 Display Unit Disassembly: étape 4, image 1 de 1
    • Disconnect the touch controller cable.

  5. GE Carescape R860 Display Unit Disassembly: étape 5, image 1 de 1
    • Disconnect the Keypad harness at the Carrier Board.

  6. GE Carescape R860 Display Unit Disassembly: étape 6, image 1 de 1
    • Disconnect the encoder harness from the Carrier Board.

  7. GE Carescape R860 Display Unit Disassembly: étape 7, image 1 de 1
    • Disconnect the speaker harness from the Carrier Board and set the speakers aside.

  8. GE Carescape R860 Display Unit Disassembly: étape 8, image 1 de 1
    • Disconnect the alarm light harness from the Carrier Board.

  9. GE Carescape R860 Display Unit Disassembly: étape 9, image 1 de 1
    • Loosen (do not remove) the five screws at the top edge of the shield just enough to raise it and slide the shield away with a Phillips #0 screwdriver.

  10. GE Carescape R860 Display Unit Disassembly: étape 10, image 1 de 1
    • Loosen the four screws along each side of the shield with a Phillips #0 screwdriver.

  11. GE Carescape R860 Display Unit Disassembly: étape 11, image 1 de 1
    • Remove the four screws at the bottom edge of the shield with a Phillips #0 screwdriver.

  12. GE Carescape R860 Display Unit Disassembly: étape 12, image 1 de 1
    • Slide the shield down to disengage the touch controller harness connector.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Andre Fisher

Membre depuis le 01/23/20

357 Réputation

2 tutoriels rédigés

Équipe

Cal Poly, Team S7-G9, Paton Spring 2020 Membre de l'équipe Cal Poly, Team S7-G9, Paton Spring 2020

CPSU-PATON-S20S7G9

3 membres

4 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 1

30 derniers jours : 9

Total : 319