Hamilton-T1 Oxygen Cells Replacement

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. Hamilton-T1 Oxygen Cells Replacement, Oxygen Cells: étape 1, image 1 de 1
    • Flip device around until the back panel is exposed.

  2. Hamilton-T1 Oxygen Cells Replacement: étape 2, image 1 de 1
    • Remove the screw in the bottom center of the panel, and then take the panel off.

  3. Hamilton-T1 Oxygen Cells Replacement: étape 3, image 1 de 1
    • Disconnect the oxygen cell by removing the tube.

    • Then twist the cell counter-clockwise and pull to remove the oxygen cell.

    • Install the new oxygen cell and put the back cover on.

    • After the cell is installed. Run the oxygen cell calibration.

    • Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.

Conclusion

After the installation and calibration of the oxygen cell. The Hamilton-T1 Ventilator is ready to operate.