Aller au contenu principal

Newport e500 Air and Oxygen Flow Sensors Disassembly

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. Newport e500 Air and Oxygen Flow Sensors Disassembly, Newport e500 Top Cover Disassembly: étape 1, image 1 de 1
    • Place device so that the top cover is facing you.

    • Open the top cover.

  2. Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 2, image 1 de 1
    • Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.

  3. Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 3, image 1 de 1
    • Remove the cover and set aside. Retain hardware for re-assembly.

  4. Newport e500 Air and Oxygen Flow Sensors Disassembly, Newport e500 Exhalation Module Disassembly: étape 4, image 1 de 1
    • Pull up and remove the flexible rubber exhalation outlet boot (push fit).

  5. Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 5, image 1 de 1
    • Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.

  6. Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 6, image 1 de 1
    • Pull the exhalation module straight up to remove it from the GDU.

  7. Newport e500 Air and Oxygen Flow Sensors Disassembly, Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 7, image 1 de 1
    • Remove the 2 screws and washers securing the manifold/mixer assembly

  8. Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 8, image 1 de 1
    • Slide the manifold/mixer assembly forward to disengage the flow sensors

  9. Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 9, image 1 de 1
    • Gently pull each flow sensor straight out from the servo valve, (the flow sensor push-fits into the servo valve port).

  10. Newport e500 Air and Oxygen Flow Sensors Disassembly: étape 10, image 1 de 1
    • Disconnect the flow sensor cable from its 10 pin connector at the underside of the flow sensor PCB and lift out the flow sensor.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Jeremy

Membre depuis le 04/07/20

589 Réputation

6 tutoriels rédigés

Équipe

Cal Poly, Team S7-G16, Paton Spring 2020 Membre de l'équipe Cal Poly, Team S7-G16, Paton Spring 2020

CPSU-PATON-S20S7G16

3 membres

13 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 1

7 derniers jours : 1

30 derniers jours : 7

Total : 166