Philips Respironics V200 Esprit Oxygen Valve Replacement

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. Philips Respironics V200 Esprit Oxygen Valve Replacement, Oxygen Valve: étape 1, image 1 de 1
    • Before replacing the oxygen valve assembly:

    • You must install the oxygen regulator assembly before reinstalling the oxygen valve assembly

    • When reinstalling the oxygen valve, make sure the oxygen regulator and oxygen flow sensor are fully installed to the oxygen valve assembly

    • Always inspect the O-rings between interconnecting parts, and replace them if they are visibly damaged or rough to the touch.

    • Always use new O-rings supplied with replacement parts

    • Lightly lubricate O-rings with Krytox GPL226 lubricant when installing, and relubricate at every reinstallation

  2. Philips Respironics V200 Esprit Oxygen Valve Replacement: étape 2, image 1 de 1
    • Remove the oxygen flow sensor

    • (Instructions found in Chapter 10.26 of the Respironics V200/Esprit Ventilator Service Manual.)

    • Careful! Do not cut the tie wrap or disconnect the O2 F/S cable from the flow sensor.

  3. Philips Respironics V200 Esprit Oxygen Valve Replacement: étape 3, image 1 de 1
    • Remove the air flow sensor

    • (Instructions found in Chapter 10.27 of the Respironics V200/Esprit Ventilator Service Manual.)

    • Careful! Do not cut the tie wrap or disconnect the AIR F/S cable from the flow sensor.

  4. Philips Respironics V200 Esprit Oxygen Valve Replacement: étape 4, image 1 de 1
    • Remove the air valve assembly

    • (Instructions found in Chapter 10.30 of the Respironics V200/Esprit Ventilator Service Manual.)

    • Disconnect the two MTR O2 cables on the oxygen valve from the mating MTR O2 cable

    • Disconnect the CROSS SOL cable from the crossover solenoid valve

    • Using a 8-mm socket with extension, remove the three bolts that hold the oxygen valve bracket

    • Slowly pull the oxygen valve assembly away from the oxygen regulator

Conclusion

To reassemble your device, follow these instructions in reverse order.

Grace Gius

Membre depuis le 05/04/20

505 Réputation

3 tutoriels rédigés

Équipe

Cal Poly, Team S1-G4, Zimmerman Spring 2020 Membre de l'équipe Cal Poly, Team S1-G4, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G4

5 membres

5 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 0

30 derniers jours : 1

Total : 259