Aller au contenu principal
Biomed
Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.
Biomed

Introduction

This is a replacement guide for the fan in a Philips Respironics V60 ventilator. The cooling fan within the ventilator prevents the system from heating up too much. So it is important to ensure that this fan is kept clean and functioning.

For more information on replacing the fan, please refer to page 9-6 in the service manual.

  1. Philips Respironics V60 Fan Replacement, shut down the ventilator: étape 1, image 1 de 2 Philips Respironics V60 Fan Replacement, shut down the ventilator: étape 1, image 2 de 2
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. Philips Respironics V60 Fan Replacement: étape 4, image 1 de 1
    • Disconnect the internal battery from the power harness connector.

  3. Philips Respironics V60 Fan Replacement, Top Cover: étape 5, image 1 de 1
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

  4. Philips Respironics V60 Fan Replacement: étape 6, image 1 de 1
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

  5. Philips Respironics V60 Fan Replacement, Fan: étape 7, image 1 de 1
    • Remove the fan mount push pins that attach the fan to the back panel.

    • Pliers may be required.

  6. Philips Respironics V60 Fan Replacement: étape 8, image 1 de 1
    • Disconnect fan cable from motor controller.

  7. Philips Respironics V60 Fan Replacement: étape 9, image 1 de 1
    • Remove fan,

    • from back panel.

  8. Philips Respironics V60 Fan Replacement: étape 10, image 1 de 1
    • Remove fan filter assembly (the fan filter and housing), from outside the back panel.

    • When installing, orient the fan with the fan cable at the top inside of the fan, and orient the fan filter and housing as shown.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Rie Kuraoka

Membre depuis le 23/01/2020

835 Réputation

7 tutoriels rédigés

Équipe

Cal Poly, Team S7-G10, Paton Spring 2020 Membre de l'équipe Cal Poly, Team S7-G10, Paton Spring 2020

CPSU-PATON-S20S7G10

2 membres

9 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 1

30 derniers jours : 9

Total : 332