Philips Respironics V60 Top Cover Replacement

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. Philips Respironics V60 Top Cover Replacement, shut down the ventilator: étape 1, image 1 de 2 Philips Respironics V60 Top Cover Replacement, shut down the ventilator: étape 1, image 2 de 2
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. Philips Respironics V60 Top Cover Replacement: étape 4, image 1 de 1
    • Disconnect the internal battery from the power harness connector.

  3. Philips Respironics V60 Top Cover Replacement, Top Cover: étape 5, image 1 de 1
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

  4. Philips Respironics V60 Top Cover Replacement: étape 6, image 1 de 1
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Rie Kuraoka

Membre depuis le 01/23/20

886 Réputation

7 tutoriels rédigés

Équipe

Cal Poly, Team S7-G10, Paton Spring 2020 Membre de l'équipe Cal Poly, Team S7-G10, Paton Spring 2020

CPSU-PATON-S20S7G10

2 membres

9 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 0

30 derniers jours : 4

Total : 266