Aller au contenu principal

ResMed AirSense 10 Air Outlet Adapter Replacement

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. ResMed AirSense 10 Air Outlet Adapter Replacement, Air Outlet Adapter: étape 1, image 1 de 2 ResMed AirSense 10 Air Outlet Adapter Replacement, Air Outlet Adapter: étape 1, image 2 de 2
    • Position the ResMed Airsense 10 with the water reservoir facing towards you.

    • Remove the ResMed water tank.

  2. ResMed AirSense 10 Air Outlet Adapter Replacement: étape 2, image 1 de 3 ResMed AirSense 10 Air Outlet Adapter Replacement: étape 2, image 2 de 3 ResMed AirSense 10 Air Outlet Adapter Replacement: étape 2, image 3 de 3
    • Use the spudger to disable the lock on the air outlet elbow.

    • Remove the air outlet elbow through the back opening on the ResMed.

    • The Air outlet elbow is now free and ready to be replaced.

Conclusion

To reassemble your device, follow these instructions in reverse order.