Viasys Avea Exhalation Corner Assembly Removal

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.
    • Check to see if the exhalation filter assembly is installed, if it is, remove the exhalation filter assembly.

    • Rotate the metal locking lever on the lower right of the ventilator body forward to an open position.

    • Remove the exhalation filter assembly from the ventilator body by pulling straight down.

    • Using a long Phillips screwdriver, remove the Phillips screw located at the top of the exhalation filter assembly well, if applicable.

    • Using a long Phillips screwdriver, remove the Phillips screw located at the exhalation port marked EXH, if applicable.

    • Using a long Phillips screwdriver, remove the Phillips screw from the outside of the plastic cover.

    • Using a long Phillips screwdriver, remove the Phillips screw on the font left wall.

    • Using a long Phillips screwdriver, remove the Phillips screw on the right side of the wall.

    • Set the plastic cover aside.

    • Using a long Phillips screwdriver, remove the 3 Phillips screws on the metal re-enforcement cover, if applicable, and set aside.

    • Remove the Exhalation Corner Assembly.

Conclusion

To reassemble your device, follow the steps in the reverse order.

Une autre personne a terminé cette réparation.

Alden Glassey

Membre depuis le 04/19/18

498 Réputation

3 tutoriels rédigés

Équipe

Cal Poly, Team S1-G1, Zimmerman Spring 2020 Membre de l'équipe Cal Poly, Team S1-G1, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G1

5 membres

6 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 3

30 derniers jours : 6

Total : 99