Introduction

Battery replacement is recommended every 300 charge cycles. Use this guide to replace the battery in your 6000 series Welch Allyn Connex Vital Signs Monitor.

PLEASE NOTE: This repair guide was developed by the iFixit team based on Welch Allyn's own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Welch Allyn.

  1. Unplug the AC power connection cable.
    • Unplug the AC power connection cable.

    Ajouter un commentaire

  2. Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.
    • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

      • Always grasp the hose by the connector. Do not pull on the hose itself.

    • Pull the connector away from the connector port.

    Ajouter un commentaire

    • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

      • Always grasp the cable by the connector. Do not pull on the cable itself.

    • Pull the connector away from the connector port.

    Ajouter un commentaire

    • Grasp the temperature probe and pull it up to remove it from the monitor.

    Ajouter un commentaire

  3. Répandez la magie de Noël
    Obtenez 12 $ de réduction sur les commandes de 50 $ ou plus avec le code FIXMAS12
    Répandez la magie de Noël
    Obtenez 12 $ de réduction sur les commandes de 50 $ ou plus avec le code FIXMAS12
    • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

    Ajouter un commentaire

    • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

    Ajouter un commentaire

    • Remove the flathead screw on the USB networking door.

    • Loosen the captive Phillips #2 screw securing the monitor to the stand.

    Ajouter un commentaire

    • Holding the monitor securely, open the USB networking door.

      • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

    Ajouter un commentaire

    • Detach any accessory USB cables from USB ports on the monitor.

    Ajouter un commentaire

    • Slide the monitor off of the stand tray.

    Ajouter un commentaire

    • Place the monitor upside down on a clean surface.

    • Insert a coin into the slot and push to open.

      • Choose a coin that fits comfortably into the slot.

    • Remove the battery cover.

    Ajouter un commentaire

    • Use the plastic label to remove the battery from its recess.

    Ajouter un commentaire

Conclusion

To reassemble your device, follow these instructions in reverse order.

4 autre(s) ont terminé cette réparation.

kyle2

Membre depuis le 03/06/2010

101 Réputation

1 tutoriel rédigé

Équipe

Team Freedom Membre de l'équipe Team Freedom

Community

1 membre

1 tutoriel rédigé

Ajouter un commentaire

Voir les statistiques:

Dernières 24 heures : 0

7 derniers jours : 2

30 derniers jours : 12

Total : 952