Aller au contenu principal

Zoll Impact Uni-Vent 731 Power Knob Kit Replacement

Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Ce dont vous avez besoin

  1. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement, Power Knob Kit: étape 1, image 1 de 1
    • Loosen but do not remove the(4) 8-32 X 3 screws on the outer air intake

    • Pg. 29

  2. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 2, image 1 de 1
    • Loosen and remove the (2) 10-32 Keps nuts and the (4) 6-32 X 2 screws. Remove the back case by lifting from the ventilator

    • Pg. 29

  3. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 3, image 1 de 1
    • Remove the (2) 4-40 X ¼ screws on the Dovetail Mounting Bracket and disconnect the ribbon cable on the PIM PCB by simultaneously applying pressure on the two locking “ears”

    • Pg. 29

  4. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 4, image 1 de 1
    • Flip the ventilator over and remove the front case assembly by lifting it straight up away from the ventilator module

    • Pg. 29

  5. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 5, image 1 de 1
    • Remove the bezel by loosening and removing the (7) 4-40 X ¼ screws that hold the bezel to the front case

    • Pg. 29

  6. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 6, image 1 de 1
    • On the membrane panel assembly, loosen the 6-32 X ¼ set screw using a 1/16 Ball Hex Driver (located in the power knob) and lift the damaged knob from the membrane panel

    • Pg. 30

  7. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 7, image 1 de 1
    • Place the replacement knob on the membrane panel switch -align with flat on switch –insert the included 6-32 X ¼ set screw and tighten

    • Pg. 30

  8. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 8, image 1 de 1
    • Install the new bezel by tightening the included (7) 4-40 X ¼ screws

    • Pg. 30

  9. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 9, image 1 de 1
    • Place the front case assembly over the vent module and tighten the (2) 4-40 X ¼ screws unto the dovetail mounting bracket. Reconnect the ribbon cable unto the PIM PCB. Make sure the two locking “ears” lock into position

    • Pg. 30

  10. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 10, image 1 de 1
    • Attach the back case to the vent module and align cover with handle, air intake housing and dovetail mounting studs. Insert and tighten the included (2) 10-32 Keps nuts and the (4) 6-32 X 2” screws

    • Pg. 30

  11. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 11, image 1 de 1
    • Tighten the (4) 8-32 X 3 screws on the outer air intake

    • Pg. 30

  12. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: étape 12, image 1 de 1
    • Perform HiPot Testing then perform Calibration and Functional Test using the RCS

    • Pg. 30

Conclusion

To reassemble your device, follow these instructions in reverse order.

Emily

Membre depuis le 05/04/20

206 Réputation

1 tutoriel rédigé

Équipe

Cal Poly, Team S1-G5, Zimmerman Spring 2020 Membre de l'équipe Cal Poly, Team S1-G5, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G5

5 membres

5 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 0

30 derniers jours : 0

Total : 157