+ | [flag|biomed] |
- | [summary]Repair guides and service information for Ohmeda Centiva 5 ventilators.[/summary] |
+ | [title|Ohmeda Centiva 5] |
+ | |
+ | [summary]Repair and service information for the Ohmeda Centiva 5 ventilator. This medical device provides general and ICU ventilation for adult and pediatric patients.[/summary] |
[summary_image|2047146] | |
[document|6908|Ohmeda Centiva_5 Service_Manual|display=list] | |
+ | |
+ | [page_title|Ohmeda Centiva 5 Repair] |
+ | |
+ | == Background and Identification == |
+ | |
+ | The Datex-Ohmeda Centiva/5 is a critical care ventilator that maintains and supports patient breathing. The device, which offers both invasive and non-invasive options, can assist adult and pediatric patients. |
+ | |
+ | Manufactured for GE Healthcare, the Centiva/5 has three major components: the main engine, control panel, and expiration valve. This software-controlled system offers volume-regulated, pressure-regulated, and spontaneous ventilation modes. |
+ | |
+ | Only physicians or qualified medical professionals should operate the Centiva/5. A maintenance check for safety and proper function is suggested every 12 months. |
+ | |
+ | The Centiva/5 is a compact ventilator that comes with two detachable display options. |
+ | |
+ | ***Dimensions for 6.5-inch Display*** |
+ | |
+ | * Height: 21 cm/8.3 in |
+ | * Width: 26 cm/10.3 in |
+ | * Depth: 29 cm/11.4 in |
+ | * Weight: 13 kg/28.7 lbs |
+ | |
+ | ***Dimensions for 12-inch Display*** |
+ | |
+ | * Height: 32 cm/12.6 in |
+ | * Width: 32 cm/12.6 in |
+ | * Depth: 32 cm/12.6 in |
+ | * Weight: 14.7 kg/32.3 lbs |
+ | |
+ | The device has flexible placement options and a backup battery. |
+ | |
+ | == Technical Specifications == |
+ | |
+ | === Physical Specifications === |
+ | |
+ | ***Centiva/5 with 6.5 inch Display*** |
+ | |
+ | * ***Footprint:*** 25 x 24 cm/9.8 x 9.4 in |
+ | * ***Height:*** 21 cm/8.3 in |
+ | * ***Width:*** 26 cm/10.3 in |
+ | * ***Depth:*** 29 cm/11.4 in |
+ | * ***Weight:*** 13 kg/28.7 lb |
+ | |
+ | ***Centiva/5 with 12 inch Display*** |
+ | |
+ | * ***Footprint:*** 25 x 24 cm/9.8 x 9.4 in |
+ | * ***Height:*** 32 cm/12.6 in |
+ | * ***Width:*** 32 cm/12.6 in |
+ | * ***Depth:*** 32 cm/12.6 in |
+ | * ***Weight:*** 14.7 kg/32.3 lb |
+ | |
+ | === Ventilator Operating Specifications === |
+ | |
+ | ***Controls and Ranges*** |
+ | |
+ | * ***FiO2:*** 21 to 100% |
+ | * ***Rate:*** 4 to 100 bpm |
+ | * ***Inspiratory/expiratory ratio:*** 4:1 to 1:9.9 VCV, SIMV-VC, 9.9:1 to 1:9.9 Bi-Level mode |
+ | * ***Tidal volume range:*** 20 to 2,000 mL |
+ | * ***Set pressure limit:*** 10 to 60 cm H2O |
+ | * ***Max. inspiratory pressure limit:*** 70 cm H2O |
+ | * ***PEEP/CPAP:*** Off to 35 cm H2O |
+ | * ***Set inspiratory flow:*** 2 to 99.9 L/min |
+ | * ***Max. inspiratory flow:*** ~180 L/min (120 L/min continuous flow) |
+ | * ***Inspiratory plateau:*** 0 to 75% of inspiration time, as a resulting parameter |
+ | * ***By-flow:*** 3 to 30 L/min |
+ | * ***Inspiratory pressure:*** 1 to 59 cm H2O |
+ | * ***Pressure Support (PS) level:*** 0 to 59 cm H2O |
+ | * ***Pressure Support Ramp:*** 0.1 to 0.5 seconds |
+ | * ***PS end flow:*** 10 to 60% of peak PS flow |
+ | * ***Trigger sensitivity:*** 0.5 to 20 L/min |
+ | * ***Trigger time window:*** 20 to 90% of expiration time |
+ | * ***Bi-Level ramp:*** 0.1 to 3 seconds |
+ | |
+ | ***Ventilation Modes*** |
+ | |
+ | * VCV, aVCV |
+ | * SIMV-VC |
+ | * Bi-Level |
+ | * Apnea Backup |
+ | * CPAP/PSV |
+ | |
+ | ***Ventilator Monitoring*** |
+ | |
+ | * ***Airway pressure:*** -9 to +90 cm H2O |
+ | * ***Patient flow:*** -150 to +200 L/min |
+ | * ***Tidal volume:*** 0 to 2,500 mL, resolution 1 mL |
+ | * ***Minute volume:*** 0 to 99 L/min, resolution 0.1 L/min |
+ | * ***Rate Spont:*** 0 to 200 breaths per minute |
+ | * ***O2 Inspiratory:*** 0 to 100% |
+ | * ***Peak pressure:*** -9 to 90 cm H2O |
+ | * ***Plateau pressure:*** -9 to 90 cm H2O |
+ | * ***Mean airway pressure:*** -9 to 90 cm H2O |
+ | * ***Pmin:*** -9 to 90 cm H2O, measured |
+ | * ***PEEP:*** -9 to 90 cm H2O |
+ | * ***Resistance:*** 1 to 99 cm H2O/L/s |
+ | * ***PEEP:*** -9 to 90 cm H2O |
+ | * ***PEEPi: -***9 to 90 cm H2O (measured during exp Pause) |
+ | * ***RSBI:*** 1 to 999 |
+ | |
+ | ***Source:*** [link|http://www.berkeley-medical.com.br/imagens/produtos/2.pdf|Ventilator Operating Specifications] |
+ | |
+ | == Additional Information == |
+ | |
+ | [link|http://www.frankshospitalworkshop.com/equipment/documents/ventilators/user_manuals/Datex-Ohmeda%20Centiva-5%20-%20User%20manual.pdf|User Reference Manual] |
+ | |
+ | [link|https://www.yumpu.com/en/document/read/10607521/centiva-5-critical-care-ventilator-mednet-healthcare-ltd|GE Healthcare Centiva/5 Critical Care Ventilator Magazine] |
+ | |
+ | [link|http://www.berkeley-medical.com.br/imagens/produtos/2.pdf|Ventilator Operating Specifications] |
+ | |
+ | [document|36588] |
+ | |
+ | [document|36606] |