Il est possible que cette traduction soit obsolète
Modifications depuis la dernière traduction
Afficher les changements
Cacher les changements
Vos contributions sont faites dans le cadre de la licence open source Creative Commons.
Accédez rapidement aux tutoriels, aux pièces et aux réponses correspondant à vos appareils
- | [flag| |
---|---|
+ | [flag|biomed] |
- | [title|VENTImotion and VENTIlogic |
+ | [title|Weinmann VENTImotion and VENTIlogic] |
[summary]Home ventilation devices for the non-invasive, non-life support ventilation of adult patients with respiratory insufficiency in whom there is evidence of autonomous respiratory drive. | |
[document|6710|VENTImotion & VENTIlogic Service and Repair Instructions|display=list] | |
+ | |
+ | [page_title|Weinmann VENTImotion and VENTIlogic Repair] |
+ | |
+ | == Troubleshooting == |
+ | |
+ | For any help identifying issues with your ventilator, check the device’s Troubleshooting Page |
+ | |
+ | [[Weinmann VENTImotion and VENTIlogic Ventilator Troubleshooting]] |
+ | |
+ | == Background and Identification == |
+ | |
+ | VENTImotion WM 24800 and VENTIlogic WM 27000 Ventilators are manufactured by Weinmann Medical Technology and were released in 2007. These devices serve as home ventilators for non-life support adult patients with respiratory insufficiency. |
+ | |
+ | This device is meant for restrictive and obstructive ventilation disorders, disorders of respiratory mechanism, central respiratory regulation disorders, neurological, muscular and neuromuscular disorders. |
+ | |
+ | Weinmann technology clearly states that the user is responsible for repairs carried out themselves and their warranty. It is advised that only Weinmann original spare parts are to be used during repair. |
+ | |
+ | If the device is to be used by a new patient, the user is to follow a series of steps to be cleaned and disinfected. Instructions and more information about the sterilization process can be found in the original manual. |
+ | |
+ | == Specifications == |
+ | |
+ | MODEL- Weinmann Hamburg VENTIlogic WM 27000 |
+ | |
+ | * Dimensions————————————--23 x 12.5 x 34 cm |
+ | * WEIGHT approx.—————————-- 4.5 kg |
+ | * Temperature range-————————-Operation: + 5 deg C to + 25 deg C |
+ | * ———————————————————Storage: - 40 deg C to + 70 deg C |
+ | * Air Pressure Range————————--600 - 1100 hPa |
+ | * Welectrical Connextion——————-115 - 230 V AC, 50-60 Hz |
+ | * ———————————————————Tolerance - 20% + 10 % |
+ | * Power Consumption————————During Operation: 230 V, 0.2 A; |
+ | * ———————————————————115 V, 0.4 A |
+ | * ——————————————————--During Stand By: 230 V, 0.043 A; |
+ | * ——————————————————--115 V, 0.068 A |
+ | * Sound Pressure Level, Alarm————Approx. 62 dB (A) |
+ | * IPAP Pressure Range———————--6 to 35 hPa |
+ | * EPAP Pressure Range———————-4 to 20 hPa |
+ | * CPAP Pressure Range———————-4 to 20 hPa |
+ | * Pressure accuracy—————————-+/- 0.6 hPa |
+ | * Increment- ————————————--0.2 hPa (1 hPa = 1 mbar = 1 cm H2O) |
+ | * Respiratory Frequency——————--6 to 45 l/min |
+ | * Accuracy—————————————---+/- 0.5 l/m |
+ | * Increment- ————————————— l/min |
+ | * I:E Inspiration Time————————--20 % to 67 % of respiratory period |
+ | * I:E Increment————————————1% |
+ | * I:E Accuracy—————————————+/- 1% |
+ | * Trigger Stage————————————Can be adjusted in 6 stages, separately |
+ | * ———————————————————inspiration and expiration. Trigger for |
+ | * ——————————————————--expiration can be switched off in ST mode. |
+ | * Pressure Increase Speed——————Can be set in 6 stages |
+ | * Pressure Drop Speed———————--Can be set in 6 stages |
+ | * Accuracy of Volume Measurement at 23 deg C: +/- 15% |
+ | |
+ | == Additional Information == |
+ | |
+ | https://www.medgadget.com/2008/04/ventilogic_promises_intelligent_home_ventilation.html |