Publication No. 7C8767 19OCT99

MODIFICATION INSTRUCTIONS for the Kodak X-Omat 3000 RA PROCESSOR Service Code: 3434 Modification No. 3 and the Kodak X-Omat 270 RA PROCESSOR Service Code: 3059 Modification No. 25

Type 2

Purpose:

To improve equipment reliability by installing:

  • new and improved QUAD POWER SUPPLY

  • FOAM STRIP to direct the heat away from the ELECTRICAL BOX

Important

Use qualified personnel to install this modification.

Service Effects:

Make the necessary changes to the ILLUSTRATED PARTS LISTS:

  • 3000 RA PROCESSOR, Publication No. 5B6336

  • 270 RA PROCESSOR, Publication No. 2B6844

Special Requirements:

None

Serial Numbers:

3000 RA PROCESSOR: 3001- 4410 270 RA PROCESSOR: All

Installation Time:

1 hour

Special Tools:

None

Parts Status:

MODIFICATION KIT 3H8289 is available from Service Parts Management at no charge until November, 2000.

Parts Requirements:

See Page 2.

HEALTH IMAGING

PLEASE NOTE

The information contained herein is based on the experience and knowledge relating to th subject matter gained by Eastman Kodak Company prior to publication.

No patent license is granted by this information.

Eastman Kodak Company reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to this information. Kodak shall not liable for any loss or damage, including consequential or special damages, resulting fro use of this information, even if loss or damage is caused by Kodak's negligence or other

This equipment includes parts and assemblies sensitive to damage from electrostatic discharge. Use caution to prevent damage during all service procedures.

MODIFICATION KIT 3H8289 includes:

Part No.DescriptionQty.Notes
3H8283FOAM STRIP - 3.8 x 3.8 x 63.5 cm (1 1/2 x 1 1/225in.) 1 3000 RA
3H9052FOAM STRIP - 3.8 x 1.9 x 71.1 cm (1 1/2 x 3/4 x28 in.)1 270 RA PROCESSOR
3H0648QUAD POWER SUPPLY
7C8767MODIFICATION INSTRUCTIONS 1

Table Caption: This table lists part numbers, descriptions, quantities, and notes for various items, including foam strips of different dimensions and two technical documents: a quad power supply and modification instructions.

Installing the New QUAD POWER SUPPLY

Figure 1 QUAD POWER SUPPLY

Installing the New FOAM STRIP

Figure 2 3000 RA PROCESSOR

H150_0232ACB

H150_0232AC

Warning

Dangerous Voltage

  • [1] De-energize the PROCESSOR.
  • [2] Raise the TOP COVER.
  • [3] Remove:
  • DRYER END PANEL
  • NON-DRIVE SIDE PANEL
  • [4] Pull out the ELECTRICAL BOX and open the COVER.
  • [5] Remove the 2 SCREWS and POWER SUPPLY COVER.
  • [6] Disconnect the 2 CONNECTORS.
  • [7] Remove the 4 SCREWS and LOCK WASHERS that hold the QUAD POWER SUPPLY.
  • [8] Remove the existing QUAD POWER SUPPLY.
  • [9] Install the new QUAD POWER SUPPLY.
  • [10] Connect the 2 CONNECTORS.
  • [11] Install the COVER on the QUAD POWER SUPPLY.
  • [12] Close the COVER and slide the ELECTRICAL BOX back into the PROCESSOR.
  • [1] If you have a 270 RA PROCESSOR, advance to Step 3.
  • [2] Remove the backing from the new 3.8 x 3.8 x 63.5 cm (1 1/2 x 1 1/2 x 25