+ | [flag|biomed] |
---|---|
+ | |
+ | [title|MEDUMAT Basic/Basic P Ventilator] |
+ | |
+ | [summary]The MEDUMAT Basic/Basic P Ventilator is an automatic (short-term) ventilator manufactured by Weinmann. |
+ | |
+ | Model numbers - |
+ | MEDUMAT Basic: WM 22600 |
+ | MEDUMAT Basic P: WM 22650[/summary] |
+ | |
[summary_image|2042977] | |
[document|6696|Medumat Basic/Basic P Service Manual|display=list] | |
- | [title| |
+ | [page_title|MEDUMAT Basic/Basic P Ventilator Repair] |
- | [summary]The Basic / Basic p is an automatic (short-term) ventilator. |
- | Model numbers WM 22600 and WM 22650[/summary] |
+ | == Troubleshooting == |
+ | If you are having issues with your MEDUMAT Basic/Basic P Ventilator, visit page 18 of the MEDUMAT Basic/Basic P Service Manual for troubleshooting steps. |
+ | == Background == |
- | MEDUMAT Basic |
+ | MEDUMAT Basic/Basic P Ventilator is an automatic (short-term) ventilator manufactured by Weinmann Medical Technology. The MEDUMAT Basic (model number WM 22600) and Basic P (model number WM 22650) are emergency automatic resuscitators used to control respiration. MEDUMAT Basic/Basic P can be used to revive patients in emergency situations or on a long-term basis. It can also be used while transporting patients within a hospital or planned long-distance. The design is to provide controlled ventilation to people of 10kg body weight and to treat respiratory arrest. It can be preset to parameters that ensure evenly balanced ventilation and can be supplied with additional modules for aspiration and oxygen inhalation. The gas used for ventilation is highly compressed medical oxygen. |
- | * is designed to provide controlled ventilation to persons of 10 kg body weight or more |
- | * is used to treat respiratory arrest |
- | * can be preset to parameters that ensure evenly balanced ventilation provided that the selected maximum ventilation pressure is not exceeded. |
- | * can be supplied with additional modules for aspiration and oxygen inhalation. (Note: MEDUMAT Basic / Basic P cannot be used as a ventilator simultaneously with these modules) |
+ | Source: [link|https://bmet.ewh.org/bitstream/handle/20.500.12091/2063/QOoIdwWZywNTtt4Y.pdf?sequence=1&isAllowed=y|Engineering World Health] |
+ | |
+ | == Identification == |
+ | |
+ | The front of the ventilator has a ventilation pressure gauge, alarm panel, air mix/no air mix switch, minute volume regulator, colour code, recommendations for ventilation settings, on/off switch, and tumbler switch. A diagram can be found on page 4 of the MEDUMAT Basic/Basic P Service Manual. The serial number of device and year of manufacture can be found on the information plate on the ventilator. |
+ | |
+ | Source: [link|https://bmet.ewh.org/bitstream/handle/20.500.12091/2063/QOoIdwWZywNTtt4Y.pdf?sequence=1&isAllowed=y|Engineering World Health] |
+ | |
+ | == Technical Data == |
+ | |
+ | * ***Product category according to 93/42/EEC:*** II b |
+ | * ***Device dimensions D x W x H:*** 190 x 110 x 90 inc. connections |
+ | * ***Weight incl. accessories:*** approx. 1.1kg |
+ | * ***Storage:*** -40°C to +70°C |
+ | * ***Control:*** timing pulse, constant volume |
+ | * ***Gas input:*** medicinal oxygen |
+ | * ***Operating pressure:*** 2.7 to 6.0 bar |
+ | * ***High-pressure gas connection:*** external thread G 3/8 |
+ | * ***Connection to ventilation hose:*** external diameter 13mm |
+ | * ***Power supply:*** maintenance-free lithium battery 3.6V; 5.2 Ah |
+ | * ***F1 fuse:*** T 500 L 250V |
+ | * ***Ventilation hose:*** spiral silicone NW 10 |
+ | |
+ | Source: [link|https://bmet.ewh.org/bitstream/handle/20.500.12091/2063/QOoIdwWZywNTtt4Y.pdf?sequence=1&isAllowed=y|Engineering World Health] |
+ | |
+ | == Additional Information == |
+ | |
+ | * [link|https://www.weinmann-emergency.com/|Weinmann Medical Technology] |
+ | * [link|https://bmet.ewh.org/bitstream/handle/20.500.12091/2063/QOoIdwWZywNTtt4Y.pdf?sequence=1&isAllowed=y|MEDUMAT Basic / Basic P Service Manual] |
+ | * [link|https://www.nhlbi.nih.gov/health/ventilator|What is a Ventilator?] |
+ | |
+ | [document|38450] |
+ | |
+ | [document|38462] |