The Water in the Humidifier Will Not Heat Up

The water in the ventilator’s humidifier remains cold even after the device has been turned on for a while.

Humidifier is defective

Test the Ventilator with another humidifier to confirm whether or not the

current one is defective. If it is defective, return it to the manufacturer

Power Board is Defective

Test to see if the power board is defective by considering these other

issues: there is no running noise and displays will not show up or when in

automatic mode, the device cannot be switched on by a breath being

taken in. If these 2 other issues are also occurring, replace the power

board (see Weinmann VENTImotion and VENTIlogic Power Board Replacement) and return it to the

manufacturer.

There are No Displays, or they are Faulty

The displays shown on the screen of the ventilator are inaccurate or nothing is showing.

Display is defective

Replace Display (see Weinmann VENTImotion and VENTIlogic Display Board Replacement)

Cable came loose or is defective

Check all cable connections in device; replace if necessary

Control Board is defective

Test to see if the control board is defective by considering these other

issues: Tolerance of therapy pressures is >0.6 hPa after 1 minute, in

automatic mode or device cannot be switched on by a breath being taken

in. If these 2 other issues are also occurring, replace the control board

(see Weinmann VENTImotion and VENTIlogic Control Board Replacement)

There is No Running Noise and the Displays Will Not Show Up

When the ventilator is turned on, the screen is black and there is no noise coming from the machine.

There is no power supply

Check connection of Power Supply cable in the device socket. Check the

source of voltage with another device (for example; a lamp). Replace

power supply cable if necessary.

The main supply unit is defective

Replace main supply unit (see VENTImotion & VENTIlogic Service and Repair Instructions , page 33)

The cable came loose or is defective

Check all cable connections in the device; replace if necessary.

The control board is defective

Test to see if the control board is defective by considering these other

issues: Tolerance of therapy pressures is >0.6 hPa after 1 minute, in

automatic mode or device cannot be switched on by a breath being taken

in. If these 2 other issues are also occurring, replace the control board

(see Weinmann VENTImotion and VENTIlogic Control Board Replacement).

The power board is defective

Test to see if the power board is defective by considering these other

issues: the humidifier will not heat up or the device cannot be switched on

by a breath being taken in. If these 2 other issues are also occurring,

replace the power board (see Weinmann VENTImotion and VENTIlogic Power Board Replacement) and

return it to the manufacturer.

Device Cannot Be Turned on using the Film Keypad

Device will not power on when the power button on the film keypad is pressed.

The ribbon cable of the film keypad has slipped out of the connection

Check whether the ribbon cable is correctly connected to the display

board.

The Fascia Film is defective

Replace fascia film (see Weinmann VENTImotion and VENTIlogic Fascia Film Replacement)

The Power Supply Cable is Not Providing Power

The power supply cable is defective

Replace the Power Supply Cable (see Weinmann VENTImotion and VENTIlogic Power Board Replacement)

Justin Ogihara

Membre depuis le 09/18/17

29 078 Réputation

52 tutoriels rédigés

Équipe

EDU Team Membre de l'équipe EDU Team

Staff

7 membres

753 tutoriels rédigés

0 commentaires

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 2

30 derniers jours : 4

Total : 154