Ohmeda Engström Display Unit Screen Replacement

Ohmeda Engström Display Unit Screen Replacement

ryan love

ryan love et 2 autres contributeurs

Dernière mise à jour le July 3, 2020

30 minutes
Modérée
Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

Replacing the Display unit Screen is a useful skill when working with the Ohmeda Engström Ventilator. Although this is not frequently needed, the display unit will have to be replaced eventually. This guide is meant to make it easy for anyone to be able to perform this basic task.

For more information on replacing the LCD display, please refer to page 9-23 (section 9.5.5) in the technical manual.

Ce dont vous avez besoin

Outils

  1. Ohmeda Engström Display Unit Screen Replacement, Display Unit Screen: étape 1, image 1 de 1
    • Turn off display unit, the power is located on the top left

  2. Ohmeda Engström Display Unit Screen Replacement: étape 2, image 1 de 1
    • Place the Display Unit face down on an anti-static pad. Make sure that the release tabs are fully depressed.

    • Loosen four screws (L) on each side of the display unit.

    • Remove the rear enclosure and you will be able to see: the fan (26), the connector panel assembly (21), the encoder assembly (12), the battery (5), the speaker (20), and the gasket(28)

  3. Ohmeda Engström Display Unit Screen Replacement: étape 3, image 1 de 1
    • Next, unplug all five wires from the display unit screen. These wires are labeled A, B, C, D, E.

    • Unscrew ten screws that are circled on the image.

  4. Ohmeda Engström Display Unit Screen Replacement: étape 4, image 1 de 1
    • Disconect the backlight harness (G)

    • Slide Grommet (H) out of the mounting plate slot

    • Remove four screws that support the mounting plate

  5. Ohmeda Engström Display Unit Screen Replacement: étape 5, image 1 de 1
    • Lift the the LCD display away from the mounting plate. Pull some of the display ribbon cable (I) to the top side of the plate. Flip the LCD over to the left of the assembly

    • Disconnect the ribbon cable (J)

    • When replacing the LCD, pull the excess ribbon cable toward the bottom side of the plate as you lower the LCD on to the plate

Conclusion

To reassemble your device, follow these instructions in reverse order.

ryan love

Membre depuis le 04/07/20

156 Réputation

1 tutoriel rédigé

Équipe

Cal Poly, Team S7-G6, Paton Spring 2020 Membre de l'équipe Cal Poly, Team S7-G6, Paton Spring 2020

CPSU-PATON-S20S7G6

3 membres

3 tutoriels rédigés

0 commentaires de tutoriel

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 3

30 derniers jours : 14

Total : 233