Aller au contenu principal
Access to Housing in the Ulco Integrus PSV

Access to Housing in the Ulco Integrus PSV

Jake Lane

Jake Lane et 3 autres contributeurs

Dernière mise à jour le July 3, 2020

8 minutes
Modérée
Matériel médical
Ni iFixit ni ce tutoriel de réparation ne sont approuvés par ou affiliés au fabricant de ce dispositif médical.

This guide will allow the user to access the Ventilator Housing, an important step in replacing the battery and the o-rings. Before you begin this process, make sure all power is disconnected so that you can be entirely safe.

For more information on accessing the housing assembly, please refer to page 23 (section 6.1) in the service manual.

NOTE: Screw types and lengths are not specified in the manuals

    • Before disassembly, make sure the front wheel breaks are on.

    • Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

  1. Access to Housing in the Ulco Integrus PSV: étape 4, image 1 de 1
    • With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

  2. Access to Housing in the Ulco Integrus PSV: étape 5, image 1 de 1
    • Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Jake Lane

Membre depuis le 04/22/19

489 Réputation

3 tutoriels rédigés

Équipe

Cal Poly, Team S7-G13, Paton Spring 2020 Membre de l'équipe Cal Poly, Team S7-G13, Paton Spring 2020

CPSU-PATON-S20S7G13

3 membres

8 tutoriels rédigés

0 commentaires de tutoriel

Ajouter un commentaire

Nombre de vues :

Dernières 24 heures : 0

7 derniers jours : 1

30 derniers jours : 5

Total : 82